Molift 4-Point Sling Bars Recall
Brand Name: Molift Product Name: Molift 4-point sling bars (Accessory to mobile hoists) Model/Catalog Number: 1430116,1430117,1430115 Produc
Updated November 17, 2025
Molift 4-point sling bars can develop a defect where the mounting bolt breaks due to material deformation, potentially preventing the device from working properly when lifting or transferring a person who cannot stand or move independently. A broken or improperly functioning lifting device poses a fall or injury risk during transfers.
Is the ETAC A/S recalled? Yes. This product was recalled by FDA on November 17, 2025. Details below.
What is recalled
Molift 4-point sling bars (model numbers 1430116, 1430117, 1430115). These are accessory bars for Molift mobile hoists, made by ETAC A/S.
Am I affected
You're affected if you own or use a Molift mobile hoist with a 4-point sling bar and the model number on your device or its documentation is 1430116, 1430117, or 1430115.
What to do right now
Stop using this sling bar immediately. Contact ETAC A/S or your supplier to report the defect and inquire about a replacement or repair. Do not attempt to use the device if the mounting bolt is broken or the joint feels stiff or restricted.
The hazard
A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.
Brands affected
- Source
- FDA
- Category
- Medical Devices
- Recall ID
- fda-Z-1311-2026
- Classification
- Class II
- Recalled
- November 17, 2025
Get free recall alerts
New Medical Devices recalls, the moment they post.
Free forever. Unsubscribe anytime. We never sell your data.
More Medical Devices recalls

SUNS International SS6291 Elevator Switch Recall
SUNS International is recalling SS6291 Solenoid Interlock Switches used in residential elevators because the switch can get stuck in a retracted position, …
SUNS SS6291 Solenoid Interlock Switches

Residential Elevators StrikeLock Door Lock Recall
Residential Elevators is recalling its StrikeLock hoistway door locking device, used on residential elevator landing doors. The defective lock can allow the …
Residential Elevators StrikeLock Hoistway Door Locking Device
BD PurPrep Povidone-Iodine Recall
BD PurPrep povidone-iodine sterile skin prep solution (25 applicators per carton) may not be sterile due to potential product contamination. Using a non-sterile …
CareFusion 213, LLC